Description
- Storage volume infusion screens for primary and secondary solutions
- Flow selection 1 to 999 mL / h in 1 mL (XL)
- Flow selection 0 to 99.9 ml in 0.1 ml / h to 999 mL to 100 / h in 1 ml (XLM)
- Battery operation
- Autotest
- Easy Setup (charging a handheld tape)
- Automatic memory retention of all previous therapy settings and fluid supply data erased by user Until
Hospira Plum XL Specifications
Physical
- Dimensions: Approximately 7.5W x 8.75D x 8.25H inches (including pole clamp)
- Weight: Approximately 7.75 pounds (with battery)
- Housing: high impact plastic
Electric
- Power requirements: 100-130 VAC, 50 to 60 Hz, less than 35 W
- Power Cable AC Cable hospital quality. 10 feet long, with transparent plug
- Fuses: 0.5 A, 250 V, slo-blo
- Battery: sealed lead-acid, rechargeable 8 V, internal to the infusion system. Accessible for easy field replacement, with color-coded wires and polarized connector
- Battery operation (XL / XLM): A new fully charged battery Provides eight hours of operation at 125 ml / h, or 1000 ml Total volume delivered, whichever first OCCURS
Ambient
- Operating Temperature: 0 ° to 40 ° C, 10% to 90% relative humidity
- Transport / Storage Temperature: -20 ° to 60 ° C
- Transport / Storage Relative Humidity: 10% to 90%
- Transport / Storage Atmospheric Pressure: 0-10,000 feet (0 to 3,000 m) or equivalent pressure